Common faults of light incubator

The light incubator is a high-precision constant temperature light simulation test equipment with refrigeration, heating and illumination control, which is widely used in research and application fields such as biopharmaceuticals, agriculture, forestry, environmental science, animal husbandry, aquatic products, etc.

light incubator chamber 

Common faults of lab light incubator:

I. The compressor does not refrigerate

1. Under the balanced mode, the compressor is always in working condition, and the refrigeration effect is good, and the compressor can be effectively maintained.

2. In case of single system operation, the compressor will not operate until the temperature drops to a certain value, and the temperature will fluctuate greatly when the door is opened or closed, resulting in frequent startup of the compressor, which is easy to shorten its service life.

II. Uneven temperature in incubator:

1. It may be caused by placing too many articles into the incubator.

2. If the light is used for a long time, the bulb is easy to heat up and the temperature will be uneven.

III. The circulating fan does not work and the circulating air inlet is frozen

Each kind of refrigeration instrument will have defrosting cycle, during which the compressor will not refrigerate. The ice on the condenser is melted by the heating element, and the melted water is likely to accumulate in the air outlet and finally cause icing.