Comprehensive Drug Stability Test Chamber

Comprehensive Drug Stability Test Chamber

LHH series medicine stability test box/general medicine stability test box Summary: Adopt balance temperature and humidity adjusting method, select imported unclosed industrial compressor, humidity sensor, touch-screen programmable controll....

 LHH series medicine stability test box/general medicine stability test box


Summary:

Adopt balance temperature and humidity adjusting method, select imported unclosed industrial compressor, humidity sensor, touch-screen programmable controller (1-99 sections), recorder etc for deploy medicine stability test box, they are stable, safe and reliable

Features:

1 adopted 304 standard of stainless-steel chamber, semicircular arcs at comers for easy cleaning

2 the major parts such as temperature and humidity controller,compressor, circulating fan adopts import product

3 unique air duct circulating design, air ducts ensures wind distribute evenly in chamber, a test hole(diameter-25mm) is placed in left side of chamber

Specifications:

model

LHH series medicine test box

LHH series general medicine stability test box

LHH-150SD

LHH-250SD

LHH-150SDP

LHH-250SDP

LHH-150GSP

LHH-250GSP

Electrical requirements

 

AC 220V 50HZ

Temperature range

0C-65℃

NO lighting:0C-65℃

lighting:10C-65℃

Temperature control

±0.5℃

Humidity range

40-95%RH

Humidity deviation

±3%RH

Illumination range

N/A

0-6000LX

controller

Special temperature and humidity digital display instrument

Imported touch-screen programmable controller

Interior dimension(mm)

500*505*600

550*520*900

500*505*600

550*520*900

500*505*705

550*520*1020

Overall dimension(mm)

630*705*1150

680*720*1540

630*705*1150

680*720*1540

630*705*1270

680*720*1660

Timing range

3(pcs)

Satisfying the ICH2003 Q1A(2)guiding principle /GMP2005 edition Chinese pharmacopoeia stability test conditions:

1 short-time test:40C±2C/75%RH±5%RH

2intermediate qualification:30c±2C/65%RH±5%RH

3 long-run test:25C±2C/60%RH±5%RH or 30C±2C/65%RH±5%RH

Execution standard:2005 edition pharmacopoeia medicine stability test guideline and relative items in GB 10586-2006